Troubleshooting Methods
Maintenance
7 - 23
Symptom
Probable Cause
Possible Remedy
Successive results
with the same sample
are not reproducible
Deteriorated chemical sample
Visually check for change. make
fresh sample, clean flowcell,
and re-analyze.
Faulty injector valve
Replace rotor seal.
Change in mobile phase
Prepare new mobile phase.
Clean flowcell. Pump new
mobile phase until stable base-
line is achieved. Re-inject
sample.
Column losing efficiency
(peak broadening or
retention time changing)
Invert and flush the column.
Replace and equilibrate with
new column.
Peaks are broad
(loss of resolution)
Column losing efficiency
Invert and flush the column.
If regeneration is unsuccessful,
replace and equilibrate with
column.
Change in mobile phase
Prepare a new mobile phase.
Clean flowcell. Pump the new
mobile phase until stable base-
line is achieved. Re-inject
sample.
Abnormal noise
Chromatographic or electric
problems
Install the “Dummy” cell
Install a power line filter.
If abnormal noise stops, check
for chromatographic problems.
If abnormal noise continues,
check the lamp or other
electrical diagnostics.
Column aging
Install a new column
Bubbles in the flowcell
Flush out the bubbles
Plumbing leaks
Check all tubing connections for
tightness
Check the flowcell and column
outlet connection
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...