![PerkinElmer 200 UV/VIS Series User Manual Download Page 90](http://html1.mh-extra.com/html/perkinelmer/200-uv-vis-series/200-uv-vis-series_user-manual_1530268090.webp)
Using Method Mode
Editing the Current Method
4 - 10
Editing Timed Events
Timed events control the relay contact closures which in turn control auxiliary instruments
(fraction collector, autosampler, etc.). You can actuate two timed events (TE1 and TE2) a
maximum of twenty times while running the method.
To edit the timed events, follow the procedure below.
From the Method Screen, press F2
to display the Method Edit Screen.
EDIT
1
Press F5
to move to the Timed Events Screen. The data field for the Time is
highlighted.
T.E.
2
METHOD EDIT SCREEN
F5
Using the numeric keys on the keypad, type in the time at which you want the first
timed event to occur.
NOTE: If you make a mistake entering numeric values in any of the fields on this
screen, press the
key and retype the number.
clear
3
TIMED EVENTS SCREEN
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...