About Sequences
Using Sequences
5 - 3
The following table describes the parts of the Sequence Screen.
The Sequence Screen displays the following softkeys:
Number
Part of the
Screen
Description
1
SET
Indicates a line in the sequence. There can be from
one to 20 lines in the sequence.
2
METH
The method number for the Set. The method must
have one or more steps.
3
Sequence Status
“LINKD” indicates that the sequence has been set
up and is ready to run. In the figure above, the
sequence status is S03.001. The “S” indicates
“Sequence.” The “03” indicates the current set, and
“001” indicates the current injection.
4
INJ
The number of injections the autosampler will
make from each vial in the set. The range is from
zero to ninety-nine.
Softkey
Description
LINK
Pressing this softkey selects the sequence for operation. When the
Run key is pressed, the sequence will be run.
DELS
Pressing this softkey deletes the entire sequence.
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...