![PerkinElmer 200 UV/VIS Series User Manual Download Page 62](http://html1.mh-extra.com/html/perkinelmer/200-uv-vis-series/200-uv-vis-series_user-manual_1530268062.webp)
Installation
Making Input/Output Connections
2 - 12
Making a Connection between Modular Cable and Twisted Pair Wires
You will need to make this connection when you are connecting the Detector to accessories
like the LC Series 200 Pump, Autosampler, or Turbochrom. The proper way to connect the
modular cable and the two wires from one of these accessories is to use the modular connec-
tor (Part No.
0999-6604)
that is supplied in the Start Up Kit.
To make a connection:
1. Refer to Figure 2-4. Cut off about 10 mm of plastic sleeve from each of the two wires.
Insert the two wires into the two center positions (labelled “R” and “G”) in the inverted
top piece.
2. Combine the top and bottom pieces. Secure them with the Phillips screw.
3. Plug the modular connector into the I/O panel connector on the Series 200 LC Autosam-
pler. Figure 2-4 shows a typical connection between the Detector and the Series 200
Pump.
4. Refer to Figures 2-5 for an example of typical I/O connections between the Detector and
other instruments making up an LC system.
TE2
This contact closure has two modes of operation. The first
mode is identical to TE1. The second mode of operation is
associated with Scanning. If this mode is chosen, when the
detector is scanning, this contact will close for the duration
of the scan.
AZ
Actuating the input from an external contact closure has the
same function as pressing the AZ key.
COMP OUT
Analog output, 1 V/AU, 0.5 V/AU, 02 V/AU (software con-
trol)
REC OUT
Analog output, 10mV full scale for any selected range.
PUMP LEAK
Optional external leak detection circuit. Contact Customer/
Service Engineering for details.
Input\Output Connector
Description
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...