Setting Configuration Options
Configuring the Detector
6 - 5
Setting Up the Lamp
You can use the Lamp Set-Up screen to set the wavelength range you want the lamp to
cover. You make the change in this screen if you do not want to accept the default value for
the installed lamp. The Detector senses the type of lamp installed and automatically dis-
plays the appropriate type. The acceptable lamp types are Deuterium or Tungsten.
The following are the default wavelength ranges:
•
190 - 360 for the Deuterium Lamp (UV)
•
360 - 700 for the Tungsten Lamp (Visible)
When you press the LAMP softkey, the Detector displays the Lamp Setup Screen. When the
Lamp Setup Screen is displayed, you can use the cursor control keys to move to the desired
field.
From the Instrument Setup Screen, press F1 to move to the Lamp Setup Screen. The
the
λ
LIMITS field is highlighted.
1
2
3
To accept the entry and move to the next field, you can use either the cursor arrow keys, or
the
key. To select another option, press the corresponding softkey.
To change wavelength range value in the
λ
LIMITS field, press the
or
softkey.
enter
360
700
To accept and save the change(s) and return to the Instrument Setup Screen, press
the
key.
To cancel changes and return to the Instrument Setup Screen press the
key.
return
quit
4
The Lamp Status field allows you to turn the lamp on or off. You can turn the lamp off while
you are viewing the Lamp Set-Up screen. The detector automatically turns on the lamp
when you exit this screen. To turn the lamp off, press the
softkey. To turn the lamp
back on, press the
softkey, or press
key.
OFF
ON
return
LAMP SET-UP SCREEN
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...