Maintenance
Troubleshooting Methods
7 - 22
Symptom
Probable Cause
Possible Remedy
Detector response is
NOT steady at any set-
ting
Loose lamp wires
Check lamp wires
Faulty lamp
Replace the lamp and align the
new lamp
Excessive line voltage
fluctuation
Check your line voltage output
Inappropriate solvent
Flush the entire system with
new solvent
Air bubbles in the solvent
inlet line
Degas the solvent
Check that solvent level is not
too low
Apply back pressure at outlet or
connect a backpressure device
(P/N 0990-7126)
Air bubbles in the flowcell
Degas the mobile phase.
Connect a backpressure valve
(P/N 0990-7126)
Bypass the column and pump
mobile phase at 5 mL/min (not
through the column) for 30 sec-
onds to dislodge air
Unstable sample
Make a fresh sample, clean the
flowcell, then re-analyze
Dirty column
Flush/regenerate column
Flowcell leaking
Check the flowcell and repair or
replace if necessary
Leak in the flowcell area
Tighten the column outlet fit-
ting if necessary
Tighten the detector waste line
for crimping and replace if nec-
essary
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...