Nitrogen Purge
In some cases (for example, in cold rooms or when operating at wavelengths of 200 nm or
lower), it is recommended that you purge the monochromator (via the purge inlet) with puri-
fied nitrogen during use. The flow of nitrogen gas will help reduce or minimize thermal
expansion/contraction of the monochromator, and, in cold rooms, prevent moisture from con-
densing inside the monochromator housing. Use only connecting tubing that is free of soften-
ers or placticizers. Purge with nitrogen at approximately 1-2 psi. (3 psi maximum). Allow 10
to 12 minutes for the monochromator to become completely filled. Continue this nitrogen
flow while the instrument is in use. Purging is particularly important when the instrument
is removed from a cold room and placed in a high-humidity room at ambient temperature.
Detector Flowcell
The standard 12
µ
L flowcell has a normal flow of 1 mL/min of 50/50 ACN/H
2
O. For addi-
tional information, refer to Chapter 7, Maintenance for the specific procedures on flowcell
maintenance.
If buffers are being used, flush the flowcell
with water to remove all buffers prior to
maintenance. Buffer may precipitate and
clog the flowcell.
Flowcell backpressure should not exceed 500
psi. Damage to the flowcell and detector
occurs at higher pressures.
Do not overtighten the flowcell connections.
Damage to the flowcell could result.
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...