Editing the Current Method
Using Method Mode
4 - 9
In the RSP/PW field, press the corresponding softkey to enter the value for the
“Smoothing” mode. This field displays either RSP or PW depending on the detec-
tor’s configuration. For more information on selecting the correct value, see the
“Detector Response Time” topic in Chapter 8, Principles.
4
In the AZ field, press the
softkey to select Autozero, or press the
No
softkey to unselect Autozero.
YES
NO
5
When you are finished entering the control parameters for Step 1, you can press the
key to go back to the Method Screen, or you can add another Step to the
method.
return
6
To add another Step to the method, press the
key, or the right arrow key.
To move from one Step to another, press the
softkey. The detector prompts
you to enter the step number. Use the numeric keypad to type the number, then
press the
key.
To save the changes to the method you can store it. For information on storing the
method, see the “Storing a Method” topic later in this chapter.
Press the
key to go back to the Method Screen.
enter
STEP
enter
return
7
When you make changes to the parameters, the method status at the top of the
screen switches to CHNGD, for “changed.”
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...