![PerkinElmer 200 UV/VIS Series User Manual Download Page 56](http://html1.mh-extra.com/html/perkinelmer/200-uv-vis-series/200-uv-vis-series_user-manual_1530268056.webp)
Installation
Making Electrical Connections
2 - 6
To set the Series 200 UV/VIS LC Detector’s operating voltage:
1. Make sure the main power switch is in the OFF position. Remove the power cord from
the module.
2. Refer to Figure 2-1. Remove the power input module cover by inserting a small flat blade
screwdriver into the notch and carefully pry off the cover.
3. Remove the voltage selector card from the power input module.
Figure 2-1.
Power Input Module (exploded view).
4. Position the number (100, 120, 220, 240) on the side of the voltage selector card so that it
is opposite the voltage selector arrow. THIS NUMBER MUST CORRESPOND TO THE
VOLTAGE FOR YOUR LOCATION. Figure 2-1 shows the voltage selector card settings
for both the 120 V and the 220 V position.
5. Reinstall the voltage selector card into the power input module. The voltage selector
arrow must line up with the corresponding hole in the module cover.
Voltage Selector
Arrow
Voltage Selector
Card
Power Input
Module Cover
Notch
Power Switch
US
E O
NL
Y W
IT
H
250
V F
US
ES
100
V
120
V
220
V
240
V
Voltage Selector
Arrow
12
0
24
0
220
100
Voltage Selector Card
Orientation for 120 Volt
Operation
Voltage
Selector
Arrow
12
0
24
0
220
100
Voltage Selector Card
Orientation for 220 Volt
Operation
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...