Configuring the Detector
Setting Configuration Options
6 - 16
Checking the Wavelength Calibration
This section describes how to run the wavelength calibration check. You can run the wave-
length calibration check when you suspect that you are getting incorrect results from the
detector. When you select
λ
CHK from the Configure menu, the Detector displays the Wave-
length Calibration Check Screen. This screen allows you to run a check on the wavelength
calibration. Before starting the check, you must replace the flowcell with a Holmium Oxide
Filter. For information on replacing the flowcell, see Chapter 7, Maintenance.