Safety and Regulatory Information
Safety and Regulatory Information
S - 17
Corrosion
The following corrosion precautions apply to the standard (stainless steel) UV/VIS and the
biocompatible (titanium) UV/VIS Detectors.
Standard
All parts of the Series 200 UV/VIS Detector that contact mobile phase are made of stainless
steel, KelF, and quartz. Some of these materials are extremely sensitive to acid chlorides. If
you have questions about your mobile phase or flush solvent, contact a PerkinElmer repre-
sentative. Refer to the table below.
Table S-1. Solvents Which May Corrode the Series 200.
Biocompatible
All parts of the Series 200 UV/VIS Detector that contact the mobile phase (from the pump)
and flush solvent are made of titanium, fluoropolymers, or glass. This makes the Series 200
UV/VIS Detector inert to the most common mobile phases that are used for biological assays.
If you have questions about your mobile phase, contact a PerkinElmer representative. Refer
to the table below.
Table S-2.
Solvents Which May Corrode the Series 200.
Aqua Regia (80% HCl, 20% HNO3)
Hydrogen Peroxide
Anhydrous Chloride
Dichloromethane
Sulfuric Acid (Conc.)
Hydrochloric Acid (20% and 37%)
Hydrofluoric Acid (20%, 50%, and 75%)
Copper Chloride
Bromine
Freon 12 (wet)
Chlorinated solvents
Hydrofluorosilicic Acid (20%)
Hydrobromic Acid (20%)
Ferric Chloride
Ferrous Chloride
Mercuric Chloride (Dilute)
Aqua Regia and sulfuric acid are known to
dissolve titanium.
Aqua Regia (80% HCl, 20% HNO3)
Hydrogen Peroxide
Dichloromethane
Anhydrous Methanol
Chlorinated solvents
Sulfuric Acid
Summary of Contents for 200 UV/VIS Series
Page 1: ...Series 200 UV VIS Detector User s Manual...
Page 23: ...Safety and Regulatory Information Warnings from the User s Manual S 21...
Page 24: ...Safety and Regulatory Information Warnings from the User s Manual S 22...
Page 50: ...Getting Started with the Detector Instrument Specifications 1 22...
Page 156: ...Maintenance Troubleshooting Methods 7 26...
Page 170: ...Enabling External Control Configuring Turbochrom with the Series 200 UV VIS Detector 9 4...
Page 176: ...Index In 6...