DMTA-10084-01EN, Rev. 2, November 2018
Pulser/Receiver Adjustments
67
—
Manual
is used to manually set the PRF value.
To adjust the PRF value with PRF Mode set to Manual
In the pulser dialog box, press the Tab key to select
PRF
, and then turn the
adjustment knob to change the PRF value.
The EPOCH 6LT is a
single-shot
instrument. This means that the instrument acquires,
measures, and draws the complete A-scan with each pulse rather than using multiple
acquisitions to construct a full waveform. The measurement rate in the EPOCH 6LT is
always equal to the PRF rate unless you are using a multiplexer.
6.3
Adjusting the Receiver
The receiver settings are accessed in the
Receiver
dialog box.
To select the receiver
1.
On the setup menu, turn the adjustment knob to select the
Receiver
(
).
2.
Press the Enter key to display the receiver dialog box (see Figure 6-4 on page 68).
NOTE
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...