DMTA-10084-01EN, Rev. 2, November 2018
Software Features and Options
185
Figure 11-27 Active AWS with D rating
11.5.2
Scanning Gain
AWS codes require that you apply a certain amount of scanning gain to the
Ref B
gain
value. This allows you to locate flaws that might be smaller or deeper in the test piece
than the reference flaw.
To add scanning gain
Use the
Gain
adjustment box to adjust the gain as necessary (see “Adjusting the
Gain (System Sensitivity)” on page 59 for instructions on using the
Gain
adjustment box).
To display a D indication rating value, the gated echo must peak at an amplitude less
than 110 % FSH (full-screen height). You may need to turn off the scanning gain, and
even make further gain adjustments, to bring the echo peak on screen.
NOTE
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...