DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
166
4.
Select the
Off
button to turn the scanning gain off completely.
Figure 11-12 on page 166 shows a
Standard
DAC setup with 6 dB of scanning
gain added.
Figure 11-12 Standard DAC with scanning gain off
When reference correction is active, the digital comparison between a captured
reflector and the DAC curve is accurate even with the scanning gain applied to the
inspection, provided that the gated echo is not saturated. Figure 11-13 on page 167
shows the same setup as above but with reference correction active. Notice that the
scanning gain has been removed from the dB-to-curve measurement in the
measurement reading 5 box. The instrument compares the echo height to the DAC
curve, compensates for the added scanning gain, and reports the true-amplitude
comparison.
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...