DMTA-10084-01EN, Rev. 2, November 2018
Chapter 6
66
P/E
Used to select pulse-echo mode, where a single element transducer sends and
receives the ultrasonic signal. You must use the red T/R transducer connector in
pulse-echo mode.
Dual
Used to select pitch-catch mode, where a dual-element transducer has one
element that transmits the ultrasonic signal and another element that receives the
signal. Use the transducer connection labeled T/R as the transmit connector.
Thru
Used to select the thru-transmission mode, where two separate transducers are
typically located on opposite sides of the test specimen. One transducer transmits
the ultrasonic signal, the other receives. Use the transducer connection labeled
T/R as the transmit connector.
To compensate for the one-way sound path in thru-transmission (
Thru
)
mode, the
EPOCH 6LT does not divide transit time by two when calculating thickness
measurements.
To select the test mode
In the pulser dialog box, press the Tab key to select
Mode
, and then turn the
adjustment knob to select and set a mode.
6.2.6
Pulse Repetition Frequency (PRF)
The EPOCH 6LT allows you to manually adjust the PRF from 10 Hz to 2000 Hz in
50 Hz (coarse) or 10 Hz (fine) increments. The instrument has an
Auto-PRF
mode to
automatically adjust the PRF based on screen range.
To select a PRF adjustment method
1.
In the pulser dialog box, press the Tab key to select
PRF Mode
.
2.
Turn the adjustment knob to select
Auto
or
Manual
.
—
Auto
is used to automatically set the PRF value based on the chosen screen
range.
NOTE
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...