DMTA-10084-01EN, Rev. 2, November 2018
Chapter 8
98
Figure 8-6 Calibration
—
Ready to collect thick echo
13. Press the Enter key to capture the thick echo.
When the calibration is complete, the following message is briefly displayed at
the bottom of the calibration dialog box:
SUCCESS!
Then the calibration dialog
box is replaced by the setup menu. On the live A-scan, measurement reading box
five displays the calibrated value of the 0.400 in. step (see Figure 8-7 on page 99).