![Olympus EPOCH 6LT User Interface Manual Download Page 142](http://html2.mh-extra.com/html/olympus/epoch-6lt/epoch-6lt_user-interface-manual_3934183142.webp)
DMTA-10084-01EN, Rev. 2, November 2018
Chapter 9
132
All measurements above this value are colored green.
All measurements between the
Low Range Thickness
value and the
High Range
Thickness
value are colored yellow.
To configure the grid size and color
1.
In the grid dialog box, press the Tab key to highlight the
Grid Size
box.
2.
Turn the adjustment knob to select
Half Size
or
Full Size
.
3.
Press the Tab key to highlight the
Display Color
box.
4.
Turn the adjustment knob to select
Mono
or
Color
.
If you select
Color
, the
Low Range Thickness
and
High Range Thickness
boxes
appear. Then complete the following steps:
a)
Press the Tab key to select
Low Range Thickness
.
b)
Turn the adjustment knob to change the low range thickness value.
c)
Press the Tab key to select
High Range Thickness
.
d)
Turn the adjustment knob to change the high range thickness value.
9.5.3
Using the Grid
Before using the grid, you must first activate it (see “Activating the Grid View” on
page 130). If you set
Grid Enable
to
On
, and then press the ESC key to dismiss the
grid
dialog box, the grid view appears on the screen, including the grid and the
Grid
controls. If you press the ESC key while in grid view, the sidebar menu replaces the
Grid
controls.
The grid controls include the arrow buttons and
Save
button.
To save data in the grid on the live A-scan display
1.
To immediately begin populating the grid with data, make a measurement, turn
the adjustment knob to highlight
Save
, and press the Enter key.
The file ID is incremented, the data is saved to that location, and a new row in the
grid is created and displayed to show the newly saved data.
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...