DMTA-10084-01EN, Rev. 2, November 2018
Software Features and Options
179
DGS/AVG curve gain adjustments are applied to both the echo height and curve
height to maintain the amplitude ratio and, therefore, the sizing comparisons.
Figure 11-23 on page 179 shows a DGS/AVG setup with curve gain in use to provide
accurate echo-amplitude measurement by placing the echo close to 80 % FSH.
Figure 11-23 Gain curve adjusted DGS
11.4.5
Registration Level Adjustment
The registration level of the DGS/AVG curve defines the height of the main curve. The
curve height represents the amplitude from a flat-bottom hole with a diameter of the
registration level at different depths. This is usually equal to the critical flaw size for
the application. The EPOCH 6LT allows you to adjust this registration level during a
live inspection.
NOTE
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...