![Olympus EPOCH 6LT User Interface Manual Download Page 211](http://html2.mh-extra.com/html/olympus/epoch-6lt/epoch-6lt_user-interface-manual_3934183211.webp)
DMTA-10084-01EN, Rev. 2, November 2018
Software Features and Options
201
Figure 11-40 Correctly triggered measurement
11.7.3.3
Echo-to-Echo Measurements
The corrosion module echo-to-echo mode measures the distance between an echo in
Gate 2 and an echo in Gate 1. Echo-to-echo measurements are most commonly used to
determine the thickness of coated materials, since the first back wall measurement
includes the thickness of the coating. Measuring the distance between the second
successive back wall and the first back wall provides a more accurate measurement of
the base material, so you can ignore the coating layer.
To make an echo-to-echo measurement
1.
In the corrosion module (
)
Corrosion
dialog box, set
Gate2
to
On
(see
Figure 11-41 on page 202).
Measurement reading 5 displays the measurement from Gate 2 to Gate 1 (2-1).
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...