![Olympus EPOCH 6LT User Interface Manual Download Page 181](http://html2.mh-extra.com/html/olympus/epoch-6lt/epoch-6lt_user-interface-manual_3934183181.webp)
DMTA-10084-01EN, Rev. 2, November 2018
Software Features and Options
171
To activate and set up the customized curves
1.
Activate the DAC/TCG feature (see “Enabling the DAC/TCG Feature” on
page 154).
2.
In the DAC/TCG dialog box, turn the adjustment knob to set the
DAC/TCG
Mode
to
Custom
.
3.
Set the
Curve Type
to either
Curve
or
Straight
(polynomial or straight-line
segments).
4.
Select the
Curve no.
to be used in addition to the main curve (see Figure 11-17 on
5.
For each warning curve, choose
Curve
<
1–6
>
dB
, and set the value compared to
the main curve.
Figure 11-17 Custom DAC setup
The
Custom
DAC setup and functionality are the same as the Standard and ASME III
modes discussed earlier in this section. Figure 11-18 on page 172 shows a completed
Custom
DAC setup.
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...