DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
164
Save
This parameter saves the DAC curve to internal memory.
Auto XX%
This parameter allows you to return to DAC acquisition mode, providing access
to the
Add
,
Delete
,
Erase
, and
Done
functions. (
Edit
is not available in the TCG
view.)
While DAC/TCG is active, you have full control of the
Range
,
Delay
, and
Zoom
settings. With this control, you can focus on areas of interest within the DAC setup.
Figure 11-10 on page 164 shows a reduced range with delay.
Figure 11-10 A small range DAC
11.3.5
Gain Adjustment Options
The DAC/TCG mode includes three separate types of gain adjustment for each
DAC/TCG setup. These gain adjustments allow for better inspection precision, easy
manipulation of curves and live peak information, and transfer correction.
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...