DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
196
Figure 11-35 Transducer Setup page
10. Press ESC twice to dismiss the
Transducer
Setup
page and display the
Do-Zero
dialog box (see Figure 11-36 on page 196).
Figure 11-36 Do-Zero dialog box
11. Press the Enter key to perform a Do-Zero and begin the calibration.
11.7.3.1
Improving Measurement Accuracy
The combination of default settings from the automatic probe recognition and the
do-zero function enable you to immediately begin measuring with a reasonable
amount of accuracy.
You can improve upon the do-zero measurement accuracy by performing a straight
beam thickness calibration while the corrosion module is active. For best results,
Olympus recommends that you perform a separate calibration in both modes
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...