DMTA-10084-01EN, Rev. 2, November 2018
Software Features and Options
161
Figure 11-7 Partial DAC curve with each echo set to 80 % FSH
11.3.3.1
Other Adjustments for Building a DAC Curve
You can increase the instrument range or increase the display delay to increase the
number of echoes that you can see.
To adjust the Range, Delay, or Zero while detecting the echo
1.
On the DAC/TCG submenu, press the ESC key to go to the sidebar menu.
2.
Turn the adjustment knob to select the
UT
button, and then press the Enter key to
display the
UT
submenu.
3.
Adjust the
Range
,
Delay
, or
Zero
parameters while detecting the echo.
To adjust the Angle, Thickness, and Curved Surface Correction values
1.
On the sidebar menu, press the Shift key, and then press the ESC key to go to the
setup menu.
2.
Turn the adjustment knob to select the
Angle
icon, and then press the Enter key to
open the Angle dialog box.
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...