DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
176
2.
Press the Enter key, and then turn the adjustment knob to adjust the
G1Start
until
the gate is over the reference indication.
3.
Press the Tab key to highlight the
Auto 80%
button.
4.
Press the Enter key to bring the reference reflector to
80 %
FSH.
5.
Press the Tab key to select
Ref
, and then press Enter to capture the reference
reflector and build the DGS/AVG curve.
Figure 11-20 Reference reflector before capture
After capturing the reference reflector, the EPOCH 6LT automatically calculates the
DGS/AVG curve(s) and displays them at the correct registration level amplitude on
the screen (see Figure 11-21 on page 177).
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...