![Olympus EPOCH 6LT User Interface Manual Download Page 172](http://html2.mh-extra.com/html/olympus/epoch-6lt/epoch-6lt_user-interface-manual_3934183172.webp)
DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
162
3.
Adjust the
Angle
,
Thick
,
X Value
, and
CSC
parameters as desired.
Figure 11-8 Completed DAC curve
11.3.4
DAC Inspection Mode
When the DAC curve is complete and the instrument is in inspection mode, the
DAC/TCG submenu displays a new set of parameters:
G1Start
The start position of gate 1 can be adjusted on the DAC/TCG submenu or the
Gate1 submenu.
DAC Gain
This adjustment enables you to manipulate the screen height/gain of both the
DAC/TCG curve and the echoes on screen. This allows for amplitude-to-curve
comparison at code compliant screen levels across the time base.
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...