DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
170
4.
Adjust the G1 Start value and use the
Add
button to add points and build the
desired DAC curve (see Figure 11-16 on page 170).
Figure 11-16 JIS curves
5.
When the curve is complete, select
Done
, then press the Enter key.
6.
Use the adjustment knob to highlight
Next DAC
and press the Enter key to cycle
through the available DAC curves.
11.3.7
Custom DAC Curves
The DAC/TCG software option features a DAC curve setup that allows you to define
up to six additional reference curves from the primary curve at varying levels from
−
24 dB to +24 dB. The
Custom
DAC mode option is ideal for unique sizing inspections
and procedure development.
Custom
DAC mode also enables you to choose between
a straight-line connection or a curved, polynomial connection of each point on the
DAC curve to meet various international or customer-specific requirements.
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...