![Olympus EPOCH 6LT User Interface Manual Download Page 89](http://html2.mh-extra.com/html/olympus/epoch-6lt/epoch-6lt_user-interface-manual_3934183089.webp)
DMTA-10084-01EN, Rev. 2, November 2018
Gates
79
7.6
Gate Tracking and Echo-to-Echo Measurements
The gate tracking feature enables you to make echo-to-echo measurements whenever
this feature is required by an application. Echo-to-echo measurements can be made
between gate 2 and gate 1.
Gate tracking maintains a constant separation between the position of the indication
in the first gate and the start position of the second gate. When gate tracking is active,
the start position value of the tracking gate (the second gate involved in the
measurement) defines the separation between the gates, not a fixed start position.
To make echo-to-echo measurements with gate 1 and gate 2
1.
Open the setup menu, and then select and open the
Gate1
icon (
). See “Setup
Mode” on page 17 for details on using the setup menu.
2.
Activate gate 1 by setting the
G1 Status
to
On
.
3.
Using the
G1 Start
parameter, turn the adjustment knob to position gate 1 over
the first echo you want to detect.
4.
Press ESC to go back to the setup menu.
5.
Select the
Gate2
icon (
), and then set
G2 Status
to
On
(see Figure 7-7 on
6.
Use the
G2 Start
parameter to position gate 2 over the second echo to detect.
The
G2 Start
position defines the separation between the position of the
indication in gate 1 and the start of gate 2.
7.
Set gate 2 as the tracking gate by setting
G2 Tracks
to
Gate1
.
The echo-to-echo mode flag (
)appears in the message bar, indicating that the
instrument is measuring the distance between positions of the indications in gate
1 and gate 2.vb
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...