DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
182
11.5.1
Enabling the AWS D1.1 Option
The first step in operating the EPOCH 6LT for AWS D1.1 inspections is to calibrate the
instrument for the transducer and test conditions. For information on the angle beam
calibration of the EPOCH 6LT, see “Calibration” on page 89, or follow the appropriate
guidelines from the American Welding Society.
To enable the AWS D1.1 option
1.
On the setup menu, select the
Sizing
icon (
) to open the
Sizing
submenu
(see “Setup Mode” on page 17 for details on icons and submenus).
2.
Select the
AWS
icon (
) to open the AWS
dialog box (see Figure 11-24 on
page 182). See “Navigating Dialog Boxes and Setup Pages” on page 19 for details
on using dialog boxes.
3.
Press the Enter key to highlight the
AWS
parameter box.
4.
Turn the adjustment knob to set the
AWS
mode to
On
.
Figure 11-24 AWS dialog box
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...