DMTA-10084-01EN, Rev. 2, November 2018
Chapter 8
90
2.
On the setup menu, select the
UT
icon (
), and then press the Enter key to
open the UT dialog box (see Figure 8-1 on page 91).
3.
Press the Enter key to select
Range
, and then turn the adjustment knob to set the
range based on the sound path range within the selected calibration block.
Use a greater range than necessary to ensure all calibration echoes appear on the
screen.
4.
Press the Tab key to move to
Velocity
, and then turn the adjustment knob to set an
approximate velocity for the test material.
Note that the
Velocity
parameter is disabled when the instrument is in time-of-
flight (µs) mode.
5.
Press the Tab key to move to
Zero
, and then adjust the zero offset of the
instrument to
0.700
(µs).
6.
Press the Tab key to move to
Delay
, then adjust the screen delay to
0.000
.
7.
Press the ESC key twice to return to the setup menu.
TIP
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...