Chapter 6
60
6.1.1
Using the Auto XX% Feature
The
Auto XX%
feature is used to quickly adjust the instrument gain (dB) setting to
bring the gated peak echo to XX % FSH (full-screen height).
Auto XX%
is especially
useful for bringing the echo from a reference indication to XX % FSH to establish the
instrument’s reference gain level (see “Setting Reference Gain and Scanning Gain” on
page 60 for further information).
The default setting for
Auto XX%
is 80 % FSH. You can adjust the FSH target value to
meet the demands of the application (see “Instrument Setup Page” on page 50).
You can use the
Auto XX%
feature to bring an echo to XX % FSH in any gate.
To use the Auto XX% feature
1.
On the live A-scan screen, press the Tab key to select the
Gain
adjustment box
(see Figure 6-1 on page 59).
2.
Press the Enter key to open the shortcut menu.
The shortcut menu opens with
Auto XX%
highlighted.
3.
Press the Enter key again to activate the
Auto XX%
feature.
Auto XX%
can be activated at any time during operation. If you do not actively select
a gate,
Auto XX%
is applied to gate 1.
6.1.2
Setting Reference Gain and Scanning Gain
Establishing the current system gain as the reference (base) level is useful for
inspections that require adding or subtracting scanning gain based on the relation to a
fixed gain level.
To set reference gain and adjust scanning gain
1.
On the live A-scan display, press the Tab key to select the
Gain
adjustment box
(see Figure 6-1 on page 59).
2.
Press the Enter key to open the shortcut menu.
3.
Turn the adjustment knob to highlight
REF GAIN
.
4.
Press the Enter key to activate the reference gain feature.
NOTE
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...