DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
174
Figure 11-19 DGS/AVG dialog box
3.
Press the Enter key to highlight the
DGS/AVG
parameter box.
4.
Turn the adjustment knob to set the
DGS/AVG Mode
to
On
.
The
DGS/AVG
dialog box enables you to define the exact probe being used for
the inspection, and to set up the DGS/AVG curve to be drawn. The adjustments
that can be made are as follows:
— Probe Type
Selects the type of probe to be used (
Straight Beam
,
Angle Beam
, or
Dual
Probe
).
— Name
Selects the actual probe to be used.
— Refl. Type
Defines the type of reflector to be used to acquire a reference indication to
build the DGS/AVG curve. For straight beam and dual probes, the available
reflectors are as follows:
○
Back wall
○
Side-drilled hole (SDH)
○
Flat-bottom hole (FBH)
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...