DMTA-10084-01EN, Rev. 2, November 2018
Software Features and Options
157
The DAC/TCG submenu should be displayed (see Figure 11-3 on page 157). If it is
not displayed, do the following:
a)
Press the Tab key to display the sidebar menu.
b)
Press the Tab key to highlight the
DAC/TCG
button, and then press the Enter
key to open the
DAC/TCG
submenu.
Figure 11-3 DAC/TCG sidebar submenu
To build a DAC curve
1.
With the DAC/TCG submenu displayed, press the Enter key to select
G1 Start
,
and then turn the adjustment knob to center gate 1 over the first echo.
2.
Press the Tab key to highlight
Add
, and then press the Enter key to add the first
point (see Figure 11-4 on page 158).
The
G1Start
button should now be highlighted.
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...