![Olympus EPOCH 6LT User Interface Manual Download Page 168](http://html2.mh-extra.com/html/olympus/epoch-6lt/epoch-6lt_user-interface-manual_3934183168.webp)
DMTA-10084-01EN, Rev. 2, November 2018
Chapter 11
158
Figure 11-4 First DAC point
3.
Turn the adjustment knob to move the gate trigger indicator to the next echo.
4.
Press the Tab key to highlight
AUTO 80%
, and then press the Enter key to bring
the second echo to 80 % full-screen height. While building the DAC curve:
— The
Delete
button deletes the most recently captured DAC point
— The
Erase
button deletes all captured DAC points (the entire curve).
5.
Press the Tab key to highlight the
Add
button, and then press the Enter key to add
the second point (see Figure 11-5 on page 159).
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...