DMTA-10084-01EN, Rev. 2, November 2018
Chapter 8
92
Figure 8-2 Angle dialog box
8.2
Calibration Modes
The EPOCH 6LT features multiple calibration modes to accurately meet the
requirements of the selected transducer, calibration test block, and application. These
calibration modes can be set in the calibration
dialog box. There are two modes for
straight beam transducers and two modes for angle beam transducers.
8.2.1
Straight Beam Modes
Straight beam calibrations can be performed using one of two modes. For the
purposes of the EPOCH 6LT calibration,
straight beam
refers to all zero-degree probes,
including contact, dual, delay line, immersion, etc. The two straight beam calibration
modes are as follows:
•
Thickness
This standard straight beam calibration mode requires you to provide two
different, known material thicknesses to properly calibrate the instrument. The
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...