![Olympus EPOCH 6LT User Interface Manual Download Page 78](http://html2.mh-extra.com/html/olympus/epoch-6lt/epoch-6lt_user-interface-manual_3934183078.webp)
DMTA-10084-01EN, Rev. 2, November 2018
Chapter 6
68
Figure 6-4 Receiver dialog box
6.3.1
Digital Receiver Filters
The EPOCH 6LT has a total instrument bandwidth of 0.2–26.5 MHz at
−
3 dB. The
instrument offers 8 standard fixed digital filters designed to improve the
signal-to-noise ratio by filtering out unwanted high and/or low frequency noise.
Because of the shifting frequency spectrum in most materials, you may need to adjust
the filter settings. In most cases, you should select either a broadband filter or a
narrow-band filter that covers the frequency of the transducer being used.
The EPOCH 6LT contains the following eight (8) standard filters.
•
0.2 MHz–1.2 MHz
•
0.2 MHz–10.0 MHz
•
0.5 MHz–4.0 MHz
•
1.5 MHz–8.5 MHz
•
2.0 MHz–21.5 MHz
•
5.0 MHz–15.0 MHz
•
8.0 MHz–26.5 MHz
Summary of Contents for EPOCH 6LT
Page 8: ...DMTA 10084 01EN Rev 2 November 2018 Table of Contents viii...
Page 10: ...DMTA 10084 01EN Rev 2 November 2018 List of Abbreviations x...
Page 16: ...DMTA 10084 01EN Rev 2 November 2018 Introduction 6...
Page 20: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 1 10...
Page 26: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 2 16...
Page 32: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 3 22...
Page 68: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 5 58 Figure 5 15 Diagnostic Test setup page...
Page 110: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 8 100...
Page 154: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 9 144...
Page 216: ...DMTA 10084 01EN Rev 2 November 2018 Chapter 11 206 Figure 11 45 Corrosion sidebar menu G2Start...
Page 220: ...DMTA 10084 01EN Rev 2 November 2018 Appendix 210...
Page 226: ...DMTA 10084 01EN Rev 2 November 2018 List of Tables 216...
Page 232: ...DMTA 10084 01EN Rev 2 November 2018 Index 222...