— SPEICH.
1.19.8 - Save parameters
SM)
only visible in service mode
2)
only for ADCs from the older Combics generation
Procedure
If the weighing platform consists of several load cells, multiply the maximum capacity
accordingly (e.g. four load cells of 50 kg each result in a maximum capacity of 200 kg)
The value must match the increment d entered under menu item [ADU.KONF]-
[STANDARD]- [EINBER] 11.4. This only applies to older ADCs!
For weighing platforms with multiple load cells: Enter the individual values for each of
the load cells (menu items [EMPFND.1] 1.19.3 … [EMPFND.4] 1.19.6 or enter the
average value across all cells (menu item [EMPFND.1] 1.19.3)
1. Open the locking switch (see Chapter
2. Activate service mode (see Chapter
3. Select the weighing platform.
4. Enter the maximum capacity of the load cell(s) in kg (menu item [NENNLST] 1.19.1).
5. Enter the resolution in the unit kg (menu item [AUFLOES] 1.19.2).
6. Enter the sensitivity of the load cell in mV/V (menu item [EMPFND.1] 1.19.3).
7. Save values for calibration without weight (menu item [SPEICH.] 1.19.8).
8. Close the locking switch.
5 Getting started
Combics 2 Series Complete Scales Models CAW2P, CAW2S
Minebea Intec
EN-57