7.14.1.1
Minimum load for application
The minimum [MIN.INIT] (3.6) amount that a component must weigh before it can be
saved in nettotal memory.
Setting: [APPLIC.] - [APPLIC.3] - [NET TOT] - [MIN.INIT] (3.6) (see chapter
Once the limit is exceeded by the load, the value can be saved.
If the load on platform is too light, the following will occur when you try to save a value:
-
The error code INF 29 is displayed
-
A warning signal is emitted (double-beep)
-
The weight value is not saved
The minimum load required for automatic taring of the container weight on the platform
(irst weight) is conigured under: [APPLIC.] - [MIN.TARE] (3.5) (see chapter
).
The minimum load can be set in 10 steps from 1 to 1000 digits (see available parameters).
The "digits" here refer to the scale intervals for the connected weighing platform. If the
interval of the connected platform is 1 g, for example, and 1000 digits are required, you
must place at least 1000 g (= 1000 intervals = 1000 digits) on the weighing platform for
initialization.
If the [APPLIC.] - [AUT.TARE] - [ON] (3.7.2) menu item is active, automatic taring is
triggered.
7.14.1.2
Component record
Setting: [APPLIC.] - [APPLIC.3] - [NET TOT] - [PRT.SAV.] (3.17) (see chapter
).
If the [PRT.SAV.] - [EACH.TIM] (3.17.2)menu item is selected, the entire component record
is printed. If the [PRT.SAV.] - [ONCE] (3.17.3) menu item is selected, the following items
are generated only once for the irst component if they have been conigured:
Blank line, dash line, date, time, ID1 through ID6, header lines 1 and 2. For subsequent
components, each "component" item (Comp xx) is followed by a blank line.
7.14.2
Net Total Formulation with Two Weighing Platforms
This mode is used for weighing large and small components at the same time. Two
Weighing Platforms This makes it possible to toggle from the small-component platform
to the largecomponent platform during measurement. Once you toggle to the large-
component platform, the
and
keys are available until a component is value is
saved. For example, you can tare a partially-illed container taken from the
smallcomponent platform on the large component platform.
The value in component memory on the small-component platform is transferred to the
large-component platform and the weight unit is converted, if necessary. The Component
and Additive display modes are both available on the largecomponent platform.
The value read by the active platform is saved in component memory. The displayed
result is accurately calculated in the active weight unit.
When you press the
key to stop a measurement series the tare memories for both
platforms are cleared, unless the large-component platform is an SBI instrument, in which
case the platform is only tared.
7.14.3
Example: Net total
Three components of a formula should be weighed.
Combics 2 Series Complete Scales Models CAW2P, CAW2S
7 Operating
EN-158
Minebea Intec