Zero setting range (menu item [ZERORNG] 1.11)
Tare/zero at power on (menu item [INIT.ZER] 1.12)
The above parameters are reset to factory settings (menu item [RESET] -
[WT.PARA] - [YES] 9.1.1) (see chapter
).
Coniguration Counter Features:
-
Counter limited to 9999
-
Counter at "P 0000" for hardware commissioning
-
Counter cannot be reset
-
Counter is updated automatically when:
-
Whenthefollowingparametersarechanged:
Menu: [SETUP] - [WP-1] - [INTERN.] - [PARAM.1] (1) (see chapter
)
Installation location (menu item [AMBIENT] 1.1)
Application ilter (menu item [ APP.FILT] 1.2)
Stability range (menu item [STAB.RNG.] 1.3)
Stability range (menu item [TARE] 1.5)
Auto zero (menu item [AUT.ZERO] 1.6)
Weight unit 1 (menu item [1.WT.UNIT] 1.7)
Menu: [SETUP] - [WP-1] - [INTERN.] - [PARAM.2] (3) (see chapter
)
Weight unit 2 (menu item [2.WT.UNIT] 3.1)
Weight unit 3 (menu item [3.WT.UNIT] 3.3)
The above parameters are reset to factory settings (menu item [RESET] -
[WT.PARA] - [YES] 9.1.1) (see chapter
)
-
The application parameters are reset to factory settings (menu item [RESET] -
[WT.PARA] - [YES] 9.1.1) (see chapter
).
7.1.11
Device parameters
-
Password protection (see Chapter
)
-
Acoustic signal (see Chapter
)
-
Keyboard (see Chapter
-
Switch of Combics automatically (see Chapter
)
-
Illumination of the display (see Chapter
-
Timer (see Chapter
-
Example 1: Switch on, zero, tare container weight, ill container, switch to gross
weight display, second weight unit or 10x resolution (see Chapter
)
.
To display both counters, press and hold the
key for longer than 2 seconds.
The "Coniguration counter" is then shown in the weight display for 3 seconds
(identiied by a [P]). Then the "Adjustment counter" is displayed for another
3 seconds (identiied by a [C]). After 6 seconds, the information display turns of
automatically.
Combics 2 Series Complete Scales Models CAW2P, CAW2S
7 Operating
EN-98
Minebea Intec