To do this, the [HEADER] setting must be activated. Then press the
key until the
desired ID entry appears, scan the barcode and save.
7.1.8
Displaying the content of the identiiers
-
key.
7.1.9
Scanning barcodes directly
Menu: [SETUP] - [BARCODE] - [INPUT] (see chapter
You can directly scan a barcode using the barcode scanner.
The barcode can contain the following codes:
-
I for write identiier
-
T for save tare memory
-
R for write reference weight
-
A for activate product data memory
Examples:
I4Anton
= write the character string to ID 4: Anton
TC1
= write 1 Kg to the preset tare memory
C
= unit: Kilograms
B
= grams
D
= carat
etc.
RC0.0023
= write 0.0023 kg as the reference weight
A1
= load product data memory 1
Menu: [SETUP] - [BARCODE] - [HEADER] (see chapter
).
The characters read from the barcode are shown in the weight value display.
7.1.10
Adjustment/Coniguration Counter for Standard Scales
Automatic recording of changes to adjustment and weighing parameters using two
independent counters. The values remain saved for the life of the component.
Adjustment Counter Features:
-
Counter limited to 9999
-
Counter at "C 0000" for hardware commissioning
-
Counter cannot be reset
-
Counter is updated automatically when:
Menu: [SETUP] - [WP-1] - [INTERN.] - [PARAM.1] (1) (see chapter
-
linearization, calibration/adjustment is successful
-
user calibration, adjustment or linearization weight is changed (menu item
[MAN.EXT.W] 1.18)
-
When the following parameters are changed:
Function of the
key (menu item [ CAL./ADJ] 1.9)
7 Operating
Combics 2 Series Complete Scales Models CAW2P, CAW2S
Minebea Intec
EN-97