Display
Cause
Remedy
INF 10
Tare not possible with assigned tare memory
Taring will only be possible again after de-
leting the application program.
INF 18
Preload too low
INF 19
Preload too high
INF 29
Minimum load not reached
Reduce minimum load (under application,
menu item [MIN.INIT] 3.6)
INF 30
BPI identiier (BPI byte) in current weighing plat-
form not deleted (COM1 is ixed to data commu-
nication XBPI
Perform factory setting of weighing para-
meters for current weighing platform
INF 31
Interface handshake activated (XOFF, CTS)
Send XON, release CTS
INF 71
Not possible to adopt the measured value (or the
entry) (e.g. control limit too small or too large)
None
INF 72
Not possible to adopt the measured value (e.g.
maximum for item counter reached)
None
INF 73
Data cannot be written or read
consult Minebea Intec customer service
INF 74
Function is locked (e.g. menu is locked, device is
already conigured on other interface)
None
NO WP
No weighing platform connected
Connecting the weighing platform
11 Error messages
Combics 2 Series Complete Scales Models CAW2P, CAW2S
Minebea Intec
EN-197