The diference between the measured value and the true weight (calibration/
adjustment weight) is displayed with the algebraic sign.
If the the procedure is canceled by pressing the
key, a protocol is printed out.
Ext. calibration
Soll + 10000 g
Diff. + 1 g
3. Place the calibration/adjustment weight onto the weighing platform.
4. Start calibration by pressing
. (Cancel calibration/adjustment by pressing
).
Once calibration is complete, the calibration/adjustment weight appears.
The GMP protocol is printed out.
7 Operating
Combics 2 Series Complete Scales Models CAW2P, CAW2S