Retracting the adjustable feet (turning clockwise) lifts the weighing platform;
extending the adjustable feet (turning counter-clockwise) lowers the weighing
platform.
Leveling the weighing platform
4.2.6
Acclimatizing the device
If a cold device is brought into a warm environment, condensation may form.
4.2.7
Connecting Peripheral Devices or Another Weighing Platform
-
Installing Connection and Interface Cables (see chapter
)
-
Preparing Cables (see chapter
)
-
Attaching the Cable Entry (see chapter
)
-
Interface Pin Assignments (see chapter
)
.
Loosen the locknuts on the adjustable feet using a wrench A/F17.
.
Extend or retract the adjustable feet using an Allen key A/F5.
.
Use the adjustable feet to align the weighing platform so that the air bubble of the
level indicator is in the center of the circle.
.
Check that all four of the adjustable feet are touching the surface.
The weight of the platform must be spread equally across the adjustable feet.
.
After the alignment, tighten the locknuts again: small platforms (1 measuring cell)
against the platform frame, large platforms (4 measuring cells) against the platform
foot.
.
Apply the load plate.
.
Keep the device disconnected from the mains and allow it to acclimatize at room
temperature for approx. two hours.
Combics 2 Series Complete Scales Models CAW2P, CAW2S
4 Device installation
EN-26
Minebea Intec