Sartorius Laboratory. 

 

L 610, L2200S, L2200P 

 

Electronic Precision Balances

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 

Installation and Operating Instructions 

 
 
 
 
 
 
 
 
 
 
 

Advanced Test Equipment Rentals

www.atecorp.com  800-404-ATEC (2832)

®

Established 1981

Summary of Contents for L 610

Page 1: ...Sartorius Laboratory L 610 L2200S L2200P Electronic Precision Balances Installation and Operating Instructions Advanced Test Equipment Rentals www atecorp com 800 404 ATEC 2832 Established 1981 ...

Page 2: ...1 Weighing pan 2 Weight display 3 ON OFF key 4 CAL key 5 Tare control 6 Menu access switch for the balance operating program 7 Leveling foot L 2200 S only 8 Power receptacle 9 level indicator L 2200 only ...

Page 3: ...ified that the equipment delivered Electronic Precision Bal ance model L 610 L 2200 S or L 2200 P is manufac tured and tested in compliance with the following DIN VDE regulations DIN IEC348 VDE0411 Safety requirements for electronic measuring apparatus DIN IEC 380 VDE 0806 Safety of electrically energized Office machines DIN IEC 601 A DE 0750 Safety of medical electrical equipment and with Article...

Page 4: ...4 Contents Equipment Supplied 5 Installation Instructions 6 Startup 7 Operation 8 Calibration 10 Balance Operating Program 11 Troubleshooting Guide 14 Accessories 15 Specifications 17 ...

Page 5: ... the following heat radiation aggressive corrosive substances vibrations draughts Do not expose the balance to extreme moisture over long peri ods Moisture in the air can condense on the surfaces of a cold balance whenever it is brought to a substantially warmer place lf you Transfer the balance to a warmer area make sure to condition it for about 2 hours at room temperature leaving it unplugged A...

Page 6: ...inted on this unit is identical to your lo cal line voltage rating Plug the cord of the AC adapter into the power receptacle of the balance Secure the connection by tightening the knurled collar Now plug the AC adapter into a wall outlet At the point of use level the balance using the leveling feet 7 so that the air bubble is entered within the circle of the level in dicator 9 only for the L 2200 ...

Page 7: ...STANDBY mode and ready to operate without warm u p POWER OFF The balance was disconnected from line power reconnection to line power after the balance was unplugged power failure CAL The calibration function has been activated In addition to grams this balance gives you a variety of other menu definable international weight unit options Select the weight unit you need from the table of the menu op...

Page 8: ...ing warmup the next time you switch it on After the power is turned on a test of all essential electronic functions is run automatically This self test ends with the readout 0 00 g it the program code setting 511 is used Now place your sample or object on the pan 1 to determine the weight Read oft the weight in the display 2 as soon as the weight unit in this case g appears as the stability symbol...

Page 9: ... 2 000 g See the Accessories Clear the pan and press the CAL key 4 for at least three sec onds until the calibration weight readout appears in the display Center the calibration weights on the pan Now the weight unit symbol is displayed An acoustic signal indicates the end of the calibration procedure ...

Page 10: ...g program With the balance turned off STANDBY state hold down the tare control 5 and briefly press the ON OFF key 3 Upon completion of the automatic self test release the tare control as soon as CH5 is displayed The Status of the bal ance operating program will appear in the weight display L Stands for the list mode In this mode you can check the menu code settings but you cannot program new menu ...

Page 11: ... symbol that appears indicates the actual setting To change any menu code settings C mode press the tare control as soon as the appropriate numeric code is displayed Brief display of BUSY and the o symbol confirms your selec tion followed by a return to zero for the 2nd digit or line How to return to the weighing mode press the tare control each time a 0 appears in the numerical sequence word line...

Page 12: ...2 2 Singapore taels tl C 1 3 1 Last decimal ON c 5 2 3 Taiwan taels tl C 1 3 2 Last decimal OFF c 5 2 4 Grains gr C 1 3 3 Last decimal at stability c 5 2 5 Pennyweights dwt C 1 3 4 All decimals at stability c 5 2 6 Mommes o c 5 2 7 Milligrams o Code Tare parameter C 5 2 8 Austrian carats o C 1 4 1 Without stabilty C 1 4 2 At stability C 5 2 0 Call Program Line Code Auto Zero C 5 0 Call Program pag...

Page 13: ...t readout changes con stantly or the special message BUSY does not go out in the weight display Unstable ambient conditions Too much vibration or the bal ance is exposed to a draft Sample does not have a stable weight Set up balance in another area Access the menu to adjust the proper code for the particular type of weighing environment Weight shows CE The Cal key 4 was not pressed when the displa...

Page 14: ...version Weight of residue in change in Statistics Net total formulation compounding Filling toward 0 Parts counting Animal weighing Calculations by a factor Calculation of the weight per unit area division Accumulator memory I D no memory Communication with Computers YDO 01 L Thermo Control infrared dryer programma ble YTC 01 L Remote display can be plugged into the interface port of the balance L...

Page 15: ...n of one of four digital filter levels Stability range d 0 25 64 selectable Ambient temperature range K 273 313 0 o C 40 o C Sensitivity drift within 283 303 K K 5 10 6 2 10 6 1 10 5 Deviation of readout when balance is tilted 1 1000 g 0 01 0 01 0 01 0 02 0 05 Pan dimensions mm 211 x 169 Balance housing W x D x H mm 220 x 235 x 75 Net weight ca kg 4 5 Power requirements voltage frequency 50 60 Hz ...

Page 16: ... Copyright by Sartorius AG Göttingen Deutschland All rights reserved No part of this publication may be printed or translated in any form or by any means without the prior written permission of Sartorius AG Sartorius AG reserves the right to make change to the technology features specification and design of the equipment without notice ...