Lock the
key so that the functions speciied above cannot be triggered:
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.JUST]- [GESPRT] (1.9.10) "Key
locked" (see also Chapter
-
Calibration and automatic or manual adjustment (not for veriied scales):
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.ABL.] (1.10) "Calibration/
adjustment procedure" (see also Chapter
).
-
Flashing
symbol as calibration alert. If several scales are connected, the
corresponding scale number also lashes:
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [ISOCAL] (1.15) "Calibration alert" (see
also Chapter
-
Release or block external adjustment:
Menu [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.EXT] (1.16) "Enable external
calibration" (see also Chapter
Example:
External calibration (see Chapter
7.2.1
Set "Mandatory calibration" function
There is a switch to change over to "Goods weighing in legal metrology" . This is located
under a cover cap to the rear left of the weighing platform housing.
Use of the veriied scale in legal metrology in the EU:
The type approval for conformity assessment is valid only for non-automatic scales; for
automatic operation with or without additional attached devices, the national regulations
applicable to the place of installation must be observed.
Externally connected IS scales:
Before you use the scale in legal metrology, use the internal calibration equipment to
calibrate the scale at the site of installation (see Chapter
).
NOTICE
Damage due to increased temperature range
The temperature range indicated on the identiication plate (°C) must not be exceeded
during operation.
Monitor the temperature during operation and cool the installation site if necessary.
For service:
External calibration for veriied scales of accuracy class
:
-
In legal metrology, "external calibration" is blocked (switch cover sealed)
-
External calibration is only possible if the seal is removed. In the event of this, the
veriication validity expires and the scale must be readjusted.
Use of the veriied scale with internal calibration equipment in legal metrology:
Opening the locking switch
.
Before use in legal metrology, perform the function [CAL.INT.] 1.9.5 "Internal
calibration" at the place of installation.
7 Operating
Combics 2 Series Complete Scales Models CAW2P, CAW2S
Minebea Intec
EN-105