The locking switch is located on the rear panel of the indicator, immediately adjacent
to the weighing platform connector.
7.2.2
Example: External calibration
External calibration and manual calibration with standard weights (weighing parameters
are factory settings).
Display [CAL.EXT.] appears for two seconds.
The prompt appears to place the calibration/adjustment weight onto the scale (here:
10,000 g).
.
Remove the cover cap.
.
Push the switch to the left (= position "open", setting for mandatory calibration).
1. Zero the scale by pressing
.
2. Start calibration of the scale by pressing and holding down the
key (e.g. the WP
symbol lashes after the calibration alert).
Combics 2 Series Complete Scales Models CAW2P, CAW2S
7 Operating
EN-106
Minebea Intec