Section 3
Getting Started
62
MikroScan 7600PRO
Operator’s Manual
Catalog 11180-94
V15.4F 050406
Mode
Details
[teMP] (deFAuLt)
Dsplays the temperature of the ponts under the color
bar. Does not, however, allow emissivity settings to be
changed or dsplayed.
[teMP & eMISS]
Dsplays the temperature and emssvty of each pont
near the pont cursor. Also allows emssvty of each
pont to be set ndvdually.
[
teMP]
Displays the specified point temperature under the color
bar. Also dsplays the temperature dfference
t
between
pont
A
and pont
b.
However, this mode does not allow
the emssvty to be corrected on an ndvdual bass.
[
teMP & eMISS]
Displays the specified point temperature under the color
bar and dsplays the emssvty settng of the pont near
the pont cursor. Also dsplays the temperature dfference
t
between pont
A
and pont
b
. However, the emissivity
s dsplayed only when the emssvty of the pont s df-
ferent from the emssvty of the whole screen.
To set the multipoint display mode:
1) Go to
[MENU]
→
[SETUP]
→
[ANALYZE]
.
Classic View Menu
[AnALyze HIGHLIGHted]
[AnALyze HIGHLIGHted]
Standard View Menu
2) Press the
jOyStICK [e]
button to gain access to the
ANALYZE SET
UP
Menu
.
[AnALyze Set uP Menu]