Section 6
Data Analysis
253
MikroScan 7600PRO
Operator’s Manual
Catalog 11180-94
V15.4F 050406
6.12.7.2 Setting the RTM Freeze Option
The 7600PRO allows you to set certan parameters for havng the
nstrument automatcally enter freeze mode whle runnng the real-
time memory. The parameters are as follows:
Freeze
Condition 1
Freeze
Condition 2
Description
OFF
OFF
Camera does not automatically enter
freeze mode while recording to RTM.
EVENT
OFF
Camera automatically enters
FREEZE mode when an event condi-
tion has been triggered. See section
6.11.1 for more information on Set-
ting the Event Condition.
OFF
MAX
FRAME
(1664)
The camera automatically enters
FREEZE mode when the maximum
number of frames held in the internal
memory has been reached.
EVENT
MAX
FRAME
(1664)
The camera automatically enters
FREEZE mode when an event condi-
tion has been triggered OR when
the maximum number of frames held
in the internal memory has been
reached. See section 6.11.1 for more
information on Setting the Event
Condition.
To set the RTM Freeze Option:
1) Go to
[MENU]
→
[MEASURE]
→
[RUN MODE]
.
Classic View Menu
[Run MOde HIGHLIGHted]
[Run MOde HIGHLIGHted]
Standard View Menu
Summary of Contents for 7600PRO