Section 6
Data Analysis
236
MikroScan 7600PRO
Operator’s Manual
Catalog 11180-94
V15.4F 050406
3) Toggle the
jOyStICK
up or down, if needed, to highlight
[EVENT]
from the list of menu choices.
[eVent HIGHLIGHted]
4) Toggle the
jOyStICK
to the right to highlight the threshold tem-
perature setting.
[Run MOde eVent tHReSHOLd
teMPeRAtuRe SettInG SeLeCted]
If a threshold value is not specified, an event condition will be
trggered accordng to the maxmum or mnmum temperature
values of the thermal mage.
5) Toggle the
jOyStICK
up or down as needed to increase or decrease
the temperature threshold value.
[Run MOde eVent tHReSHOLd
teMPeRAtuRe SettInG CHAnGed]
NOTE
If a threshold value is not
specified, the alarm will be
triggered according to the
maximum or minimum tem-
perature values of a specific
box or the whole image as
specified through the box
range options.
Summary of Contents for 7600PRO