CAN Triggering and Serial Decode
25
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
279
To display CAN serial decode, see
CAN Serial Decode
To set up the oscilloscope to capture CAN signals, see
For CAN triggering set up see
To set up CAN serial decode:
1
Press the
[Analyze]
key.
2
Press
Features
; then, select
Serial Bus
.
3
Press
Features
again (or push the Entry knob) to enable the serial bus display.
4
Press the
Mode
softkey; then, select
CAN
.
5
If the oscilloscope is stopped, press the
[Run/Stop]
key to
acquire and decode data.
You can use the horizontal
Zoom
window for easier navigation of the decoded data.
NOTE
If the setup does not produce a stable trigger, the CAN signal may be slow enough that the
oscilloscope is auto-triggering. Press the
[Trigger]
key, then press the
Mode
softkey to set the
trigger mode from
Auto
to
Normal
.