240
Keysight InfiniiVision 1200 X-Series and EDUX1052A/G Oscilloscopes User's Guide
22
Utility Settings
•
Normal
— to make the oscilloscope switch to Normal acquisition mode
whenever the
[Auto Scale]
key is pressed. This is the default mode.
•
Preserve
— to make the oscilloscope remain in the acquisition mode you
have chosen when the
[Auto Scale]
key is pressed.
Setting the Oscilloscope's Clock
The Clock menu lets you set the current date and time of day (24-hour format).
This time/date stamp will appear on hardcopy prints and directory information on
the USB mass storage device.
To set the date and time, or to view the current date and time:
1
Press
[Utility] > Options > Clock
.
2
Press the
Year
,
Month
,
Day
,
Hour
or
Minute
softkey; then, rotate
the Entry knob to set to the appropriate number.
3
Press the
Time Zone
softkey; then, rotate the Entry knob to
select your time zone.
Because licenses are based on Coordinated Universal Time
(UTC), the proper time zone setting allows licenses to work
immediately after they are generated.
The hours are shown in the 24-hour format. So 1:00 PM is hour
13.
The real-time clock allows selection of valid dates only. If a day
is selected and the month or year is changed so the day is
invalid, the day is automatically adjusted.
Setting the Gen Out Source
On G-suffix oscilloscope models (that have a built-in waveform generator), you
can choose the source of the Gen Out connector on the front panel of the
oscilloscope: